Hãng sản xuất: AppliTek / Hach
Model: EZ1011.99001C02
Xuất xứ: Bỉ
Alarm: 1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Ambient Temperature: 10 – 30 °C ± 4 °C deviation (50 – 86 °F ± 7.2 °F deviation) at 5 – 95% relative humidity (non-condensing)
Calibration: Automatic, 2-point; frequency freely programmable
Certifications: CE compliant / UL certified
Cycle Time: 10 min (dilution + 5 min)
Demineralized Water: For rinsing / dilution
Digital Outputs: Modbus RS485
Dimensions (H x W x D): 690 mm x 465 mm x 330 mm
Lower Limit of Detection (LOD): ≤ 10 µg/L
Material: Hinged part: Thermoform ABS, door: plexiglass
Wall section: Galvanised steel, powder coated
Measurement Method: Colorimetric measurement at 480 nm using bathocuproine method; conform respectively with USEPA Method 8506 and standard method APHA 3500-Cu (C)
Number of sample streams: 1 stream
Output:Modbus RS485
Power: 110 – 220 VAC, 2 A, 50/60Hz
Max. power consumption: 150 VA
Other voltages available on request
Precision: Better than 2% full scale range for standard test solutions
Protection Class: Analyser cabinet: IP55 / Panel PC: IP65
Range: 0.05 – 5 mg/L Cu(II)
Reagent Requirements: Keep between 10 – 30 °C (50 – 86 °F)
Sample Flow Rate: 100 – 300 mL/min
Sample Pressure: By external overflow vessel
Sample Quality: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Sample Temperature: 10 - 30 °C (50 - 86 °F)
Validation: Automatic; frequency freely programmable
Warranty: 1 year
Weight: 25 kg (55 lbs.)